Special features
---------PC integrated mainframeRobust and portable
Three slots for acquisition cardsAccepts any mix of GEN DAQacquisition cards
Up to 96 analog channels
200 MB/s continuous streamingSynchronized recording usingtwo GEN3i or GEN2i mainframesRemote use from external PCPerception Standard softwarewith Windows based userinterface for advanced reviewand analysis
The GEN3i is a versatile portable data
recorder. In addition, it provides all the featuresexpected from a transient recorder. The
hardware combines a full-featured, low-power,Intel Core™ i5 3rd generation Windows® PCwith a large, high-resolution, touch screen anda robust three-slot acquisition unit. The GEN3ihas the same form factor as its predecessorGEN2i, now offering more channels and moreprocessing power. The GEN3i comes with fivedifferent Windows® languages and eightdifferent Perception languages pre-installed.
Designed for operation in the field as well as inthe laboratory the GEN3i features a unique,Instrument Panel touch interface, with one-touch access to all features for daily operation.In addition, the GEN3i includes PerceptionStandard for post-processing. With a singletouch the data recorder turns into a dedicatedinstrument for analysis and sophisticatedreporting using the extra software options.
B3762-2.0 en
Data sheetGEN seriesGEN3i
GEN3i Block Diagram
Figure 1.1: Block Diagram GEN3i Portable Recorder
Windows® PC
MemoryProcessorEthernet
8 GB; DDR3 RAM
Intel 3610ME, Core™ i5 3rd generation; 2 Core, 4 threads; 2.7 GHz, 3.3 GHz turbo3 * RJ45 Ethernet connection; 1 Gbit/s
1 * RJ45 Ethernet connection with IEEE588 (PTP) support; 1 Gbit/s
1 * SFP based optical Ethernet connection with IEEE588 (PTP) support; 1 Gbit/s; supports850 and 1310 nm SFP modules.Supported on all Ethernet ports
Embedded 801.11b/g/n; , 100 and 300 Mbit/s; Wireless LAN can be hardware disabledUSB 2.0, 2 on back + 2 on front
USB 3.0, 4 on back (Using selective SSD > 100 MB/s continuous streaming)Solid State Drive (SSD), unformatted size 480 GB, 200 MB/s continuous streamingTFT SXGA touch screen, 17” / 1280x1024 resolution
2 * Display port and 1 * DVI-D connector; CRT 2048 x 1536 and DVI-D 1600 x 12003; clone mode and extended modeInternal speaker/jack plug 3.5 mmJack plug 3.5 mm
USB Keyboard and USB optical mouse
4; Direct recording control Start/Stop/Pause/Trigger
Wake on LANWireless LANUSB connectorsInternal storage PC diskDisplay
Video connectionMultiple monitors supportSpeaker/Speaker OutMicrophoneAccessoriesFront panel controls
Software
Instrument panel / Touch interface(Fully touch-optimized)DAQ softwareDAQ software options
DAQ Software and Instrument panel languagesOperating system
Operating system installed languages
Setup of instrument, Acquisition control, Display data: live / review, Basic measurements,Export and archiving, Basic reporting
Perception Standard package. Refer to Perception specification sheet for details.Analysis, Advanced Report, Video Playback, Multi Workbooks, Information, Basic FFT,Sensor Database and more
English, German, French, Chinese, Japanese, Korean, Russian, Portuguese (Brazilian)Microsoft Windows® 7 Ultimate
English, German, French, Chinese, Japanese
Other languages can be downloaded and installed using \"Windows® Update\"
HBM2B3762-2.0 en
Application Programmers Interfaces (API)
PNRF reader (free of charge)FunctionsCOM interface
Read PNRF, NRF and LRF recording files directly in your own application
The PNRF reader comes as COM interface and can be used from any application orprogramming language which supports COM automation
PNRF Software Development Kit (SDK)Matlab® integrationLabVIEW™ integrationFunctionsCOM interface
Available basic commandsExamples (free of charge)
LabVIEW™ integration (free of charge)CSI (Customer Software Interface)Functions
Installs PNRF dll’s and supplies Visual Basic, C# and C++ getting started examplesMatlab® PNRF reader install and example available within the PNRF SDKAvailable directly from National Instruments
Control Perception software from an external computer/application on Windows®, Linux,Unix, Mac OS X
All RPC commands have a COM wrapper for easier Windows® software integrationLoad and Save Perception setup files, Setup Recording, set and review Hardware Settings,Start/Stop/Pause/Trigger, monitor Live data
C++ and C# getting started example programs supplied for Windows®, source codeincluded. Unsupported Linux getting started example on request only.LabVIEW™ getting started example using RPC/COM available
Create software extension inside the Perception software by adding CSI user sheets,custom automation and extended analysis functions. Basic Windows C# sheet templateincluded. Available for all Microsoft .NET® 4 supporting languages.
Access to every Perception part: Start/Stop/Pause and Trigger, Start Manager, AcquisitionSystem, Hardware Settings, Displays, Meters, User Tables, Formulas, Calculations, DataManager, Data Sources, User variables, Notifications, Logging, Conversion Functions,Automation Actions, Sheet Manager and more, to create a dedicated application GUI thathides the entire Perception standard GUI
C# getting started example programs supplied, source code included
DCE/RPC (Distributed Computing Environment/Remote Procedure Calls)
Available basic controls & commands
Examples (free of charge)Training/Support Program
HBM offers paid professional training and support programs on all API interfaces (PNRF reader, RPC and CSI). Training program will be C#based, on-site or at central HBM location. On-site training can be customer specific. Support can be the development of a full custom softwareapplication or answering questions of software engineers.
Acquisition System
System Timebase and synchronizationCentral timebase for all acquisition cards
Accuracy
Base
Synchronization sources
± 3.5 ppm; aging after 10 years ± 10 ppmBinary, Decimal or External
IEEE1588:2008 (PTP V2: Precision Time Protocol; using 2 step master protocol)Master/Slave synchronization; Slave or single Master modeMultiple Master mode scheduled for Q3 2014IRIG and GPS scheduled for Q3 2014
Synchronization accuracy reduces as network traffic increases on Ethernet port used forPTP synchronization
± 150 ns; < 10% load no Ethernet switch used± 500 ns; < 50% load no Ethernet switch used± 5 μs; < 80% load no Ethernet switch used
To avoid excessive network traffic blocking the PTP synchronization, separate thecommunication and PTP synchronization on two independent networks.
Use 2 step master PTP aware switches to obtain best results across switches. Actual
accuracy depends on switch used. The use of switches without PTP support will create anunpredictable accuracy.
PTP synchronization accuracy
Acquisition Slots
Unused slots must be covered using the GEN DAQ blank panel. This closes the mainframe front panels for EMC/EMI and safety compliancebut also regulates the internal airflow for correct cooling of the acquisition system.
Maximum slotsAcquisition boards
Digital Event/Timer/Counter connector
3
All slots support any combination of GEN DAQ Acquisition boards1; Connected to slot A and B
B3762-2.0 en3HBM
Acquisition System
Thermal control
Every acquisition board and the acquisition system monitors its own temperatures andstatus. This is used to regulate fan speeds and reduce noise while optimizing airflow andpower consumption.
Any changes to the acquisition system configuration may change its internal thermal
gradients. As accurate calibration relies on a steady and repeatable thermal environment,calibration will be void if changes are made in the configuration. See the individual cardspecifications for calibration impact.
Calibration
GEN3i Connection Overview
Figure 1.2: GEN3i Connection Overview
HBM4B3762-2.0 en
Master/Slave Synchronization
Mainframe to mainframe phase shiftLED signalingMaster modeSlave mode
Basic synchronization
First sample
Synchronized timebaseChannel trigger exchange
Extended synchronization
Maximum number of mainframesSynchronous recording actionsSynchronous manual trigger
Compatibility
2 (Can be used in mix with GEN2i, GEN3i and GEN3t mainframes)
Start/Stop and Pause of a recording across multiple mainframes each controlled by aseparate Perception. Stop recording is a non synchronous actionUser action within Perception to trigger all mainframes synchronously
Basic synchronization features are backward compatible with GEN Series Master/Slavecard option for both Master and Slave modes. Extended synchronization features are notsupported by the Master/Slave card option. When using a mixed setup the system willautomatically work with basic synchronization only
Synchronizes the first sample in the recording for each mainframePrevents frequency drift of the sample rates within each mainframe
Synchronously exchanges every channel trigger connected to the Master/Slave trigger busto/from each connected mainframe± 100 ns
Optical link synchronized, not connected, function disabled
Basic and extended synchronization supported; 1 Slave maximum. Multiple slave supportedby Multi Slave option (Available Q3 2014)Basic and extended synchronization supported
Probe Calibration
PinsSignal
Signal amplitude
2; Signal and ground~1 kHz square wave0 V to 2 V using 1 MΩ load0 V to 1 V using 50 Ω load
B3762-2.0 en5HBM
I/O Connector
Figure 1.3: Pin assignment I/O Connector
Connector typeMating connector typeBreakout cable (included)Cable typeConnector typeLength
External Timebase In
Pulse width
Maximum frequency
Active edge
Rounding resolution
Coax
6; BNC female0.5 m (1.6 feet)
TTL compatible, Low -30 V to 0.7 V, High 2 V to 30 V100 ns min.5 MHzRising
4.01 µs; 250 kS/s and 20 kS/s acquisition cards1.01 µs; 1 MS/s and 200 kS/s acquisition cards510 ns; 2 MS/s and 200 kS/s (GN611) acquisition cards60 ns; 100 MS/s and 25 MS/s acquisition cards
Input to sample moment delayInput overvoltage protection
External Trigger In
Resolution
Minimum pulse width
Active edge
Input overvoltage protection
Delay(1)
Send to External Trigger OutTop Dead Center Rotational input
External Trigger Out
Active levelPulse width
Output impedanceShort circuit protected
Delay(1)
350 – 400 ns, plus maximum 1 full \"rounding resolution\"± 30 V DC
TTL compatible, Low -30 V to 0.7 V, High 2 V to 30 V50 ns500 ns
Selectable rising or falling± 30 V DC
± 1 μs + maximum 1 sample period (for decimal and binary time base)User can select to forward External Trigger In to the External Trigger Out BNCUsed to indicate top dead center in rotational external timebaseTTL compatible; 0 V < Low < 0.6 V; 2 V < High < 5 VSelectable High/Low/Hold High
High or Low selected: 12.8 µs
Hold High selected : Active from first trigger to end of recording49.9 Ω ± 1 %Continuous
516 ± 1 μs + maximum 1 sample period when Clock base: decimal, Filter: wideband(2)504 ± 1 μs + maximum 1 sample period when Clock base: binary, Filter: wideband(2)
(1)(2)
Delays are equal for all acquisition cards.
If analog and/or digital filter is used extra delay will be added depending on type of filter and signal frequency.
Tyco: 2-5747706-0 (D-sub, 9 pin female)Tyco: 5-747904-5
HBM6B3762-2.0 en
I/O Connector
External Event Out
FunctionActive levelPulse width
Output impedanceShort circuit protected
Delay(1)
External Start In
Minimum Pulse width
Active edge
Start response time
Input overvoltage protection
External Stop In
Minimum Pulse width
Active edge
Stop response time
Input overvoltage protection
(1)(2)
Delays are equal for all acquisition cards.
If analog and/or digital filter is used extra delay will be added depending on type of filter and signal frequency.
TTL compatible; 0 V < Low < 0.6 V; 2 V < High < 5 VSelectable Alarm or Recording Active outputSelectable High/Low for Alarm outputRecording active High output
Alarm: Active from start of alarm condition until condition endsRecording: Active until recording stops49.9 Ω ± 1 %Continuous
515 ± 1 μs + maximum 1 sample period when Clock base: decimal, Filter: wideband(2)503 ± 1 μs + maximum 1 sample period when Clock base: binary, Filter: wideband(2)TTL compatible, Low -30 V to 0.7 V, High 2 V to 30 V200 ns
Rising/falling edge; software selectableTypically 1 s when system is completely idle± 30 V DC
TTL compatible, Low -30 V to 0.7 V, High 2 V to 30 V200 ns
Rising/falling edge; software selectable
Typically 1 s when system is recording without automation± 30 V DC
Digital Event/Timer/Counter
Figure 1.4: Digital Event/Timer/Counter block diagram
Supported CardsNumber of ConnectorsConnector TypeCable Connector TypeOutput Power
Voltage
Maximum current
Event Inputs
Number of events
Levels
16 per card, 2 cards per connector
TTL Compatible, Low -30 V to 0.7 V, High 2 V to 30 V5 ± 0.5 V DC0.5 A
See specifications of acquisition cards1
44 pin, female D-type connector, AMP HD-22 series (Tyco: 5748482-5)44 pin, male D-type connector, HDP-22 series (Tyco: 1658680-1)
B3762-2.0 en7HBM
Digital Event/Timer/Counter
Figure 1.5: Logic threshold voltage levels
Overvoltage protection
Timer/Counter
Number of channels
Functions
Outputs
Number of outputs
FunctionsOutput levelsOutput resistanceOutput current
Two per card, two cards per connector
See specifications of acquisition cards that support these outputsTTL compatible; 0 V < Low < 0.6V; 2 V < High < 5 V49.9 Ω ± 1 %
20 mA, short circuit protected
Two per card, two cards per connector
See specifications of acquisition cards that support these inputs± 30 V DC
Digital Event/Timer/Counter Connector Pin Assignment
Figure 1.6: Pin diagram for Digital Event/Timer/Counter connector
QuantumX Power Supply
Connector typeMating connector typeConnector pinningOutput PowerOutput Voltage
Maximum Output Current
ODU, G81L0C-P08LFG0-0000ODU, SX1LOC−P08MFG0−0000Only GND and PWR signal connected15 Watt
> 11 V; Typical 11.5 V to 12 V
1.4 A; Current limited and short circuit protected
HBM8B3762-2.0 en
Power
Power Inlet
Total Power of unit (maximum)
47-63 Hz, 100-240 V AC250 VA, 300 VA peak
Physical, Weight and Dimensions
Weight
Mainframe
Dimensions
Height/Height with handle
WidthDepth
Acoustic NoiseTemperature SensorsCooling FansHandleTilting FeetGroundingCasingAccessories
34.2 cm/39.2 cm (13.5\"/15.4\")43.6 cm (17.2\")18.6 cm (7.3\")
The total A-weighted SPL 55 dBA @ 0.6 m maximumTemperature for monitoring and air flow control2
One carrying handle, also used for higher tilt anglesTwo retractable feet for small tilt angles4 mm Banana plugAluminum/Plastic cover
Soft carry case with strap for transportation included with hardened front and back forprotection, and storage pouches for mouse and keyboard9 kg (20.9 lbs), add ≈ 1 kg (2.2 lbs) per acquisition board installed
Figure 1.7: GEN3i Dimensions
B3762-2.0 en9HBM
Environmental Specifications
Temperature Range
Operational
Non-operational (Storage)
Thermal protection
Relative humidityProtection classAltitude
Shock: IEC 60068-2-27
OperationalNon-operational
Vibration: IEC 60068-2-34
OperationalNon-operational
Operational Environmental Tests
Cold test IEC60068-2-1 Test AdDry heat test IEC-60068-2-2 Test BdDamp heat test IEC60068-2-3 Test Ca
Non-Operational (Storage) Environmental Tests
Cold test IEC-60068-2-1 Test AbDry heat test IEC-60068-2-2 Test Bb
Change of temperature test
IEC60068-2-14 Test NaDamp heat cyclic test
IEC60068-2-30 Test Db variant 1
-25 °C (-13 °F) for 72 hours
+70 °C (+158 °F) humidity <50 % RH for 96 hours-25 °C to +70 °C (-13 °F to +158 °F)
5 cycles, rate 2 to 3 minutes, dwell time 3 hours+25 °C/+40 °C (+77 °F/+104 °F), humidity >95/90 % RH6 Cycles, cycle duration 24 hours-5 °C (+23 °F) for 2 hours+40 °C (+104 °F) for 2 hours
+40 °C (+104 °F), humidity >93 % RH for 4 days1 g RMS, ½ h; 3-axis, random 5 to 500 Hz2 g RMS, 1 h; 3-axis, random 5 to 500 Hz
Half-sine 10 g/11 ms; 3-axis, 1000 shocks in positive and negative directionHalf-sine 25 g/6 ms; 3-axis, 3 shocks in positive and negative direction0 °C to +40 °C (+32 °F to +104 °F)-25 °C to +70 °C (-13 °F to +158 °F)
Automatic thermal shutdown at 85 °C (+185 °F) internal temperatureUser warning notifications at 75 °C (+167 °F)0 % to 80 %; non-condensing; operationalIP20
Maximum 2000 m (6562 feet); operational
Harmonized Standards for CE Compliance, according to the following directives
Low voltage directive (LVD): 2006/95/EC
Electromagnetic compatibility directive (EMC): 2004/108/ECElectrical SafetyEN 61010-1 (2010)EN 61010-2-030 (2010)EN 61326-1 (2006)EMISSIONEN 55011
Industrial, scientific and medical equipment - Radio-frequency disturbance characteristics - Limits and methods ofmeasurement
Conducted disturbance: class B; Radiated disturbance: class ALimits for harmonic current emissions: class D
Limitation of voltage changes, voltage fluctuations and flicker in public low-voltage supply systemsElectrostatic discharge immunity test (ESD);
contact discharge ± 4 kV/air discharge ± 8 kV: performance criteria BRadiated, radio-frequency, electromagnetic field immunity test;80 to 2700 MHz using 10 V/m, 1000 Hz AM: performance criteria A
Electrical fast transient/burst immunity test
Mains ± 2 kV using coupling network. Channel ± 2 kV using capacitive clamp: performance criteria B
Surge immunity test
Mains ± 0.5 kV/± 1 kV Line-Line and ± 0.5 kV/± 1 kV/± 2 kV Line-earth Channel ± 0.5 kV/± 1 kV using couplingnetwork: performance criteria B
Immunity to conducted disturbances, induced by radio-frequency fields
0.15 to 80 MHz, 1000 Hz AM; 10 V RMS @ mains, 3 V RMS @ channel, both using clamp: performance criteria AVoltage dips, short interruptions and voltage variations immunity testsDips: performance criteria A; Interruptions: performance criteria C
Safety requirements for electrical equipment for measurement, control, and laboratory use - General requirementsParticular requirements for testing and measuring circuits
Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements
Electromagnetic Compatibility
EN 61000-3-2EN 61000-3-3IMMUNITYEN 61000-4-2EN 61000-4-3EN 61000-4-4EN 61000-4-5
EN 61000-4-6EN 61000-4-11
HBM10B3762-2.0 en
IRIG, IRIG/GPS (options, to be ordered separately; available Q2 2014)
IRIG(1)
Supported by IRIG and IRG/GPS optionTime Code Translator (Input)
Time Code formatsModulation ratioInput amplitudeInput impedance
Time Code Output
Time Code formatModulation ratioOutput amplitude
DC level shift
AM modulated input/output connectorsDCLS connector
Time synchronization accuracyGEN DAQ series functions
IRIG B, IEEE 1344 compliant3:1
4 V Peak-to-Peak (fixed) into 50 ΩTTL/CMOS
2 SMB sockets; one for input and one for output
Micro DP, 15-pin; some signals internally linked to Interface/Controller Module<5 µs modulated, <1 µs (DCLS)
Capture start of recording time
Synchronize Master Time Base oscillator frequency
No recording active
Recording or pause active
User notifications while recordingShort term tracking stabilityLong term tracking “Fly-wheeling”GPS(1)
Only supported by IRG/GPS optionGPS connectorGPS antennaGPS antenna cable
Time synchronization accuracyGEN DAQ series functionsGPS localization time
No recording active
Recording or pause active
User notifications while recordingShort term tracking stabilityLong term tracking “Fly-wheeling”(1)
Requires factory installation
Micro DP, 9-pin1; included
50 m (1 feet); included<1 µs
Capture start of recording time
Synchronize Master Time Base oscillator frequency2 to 15 minutes1 to 10 minutes
1 to 10 minutes plus 25 s per ms recording time deviation from IRIG timeTime marks on GPS satellites lost/restored and GPS time synchronized5.0 E-85.0 E-71 to 5 minutes
1 to 5 minutes plus 25 s per ms recording time deviation from IRIG timeTime marks on IRIG signal lost/restored and IRIG time synchronized5.0 E-85.0 E-7
IRIG A and IRIG B, IEEE 1344 compliantAM Modulated or DC level shift (DCLS)3:1 to 6:1
500 mV to 5 V Peak-to-Peak>10 kΩ
Time required to full synchronization after IRIG signal detected
Time required to full synchronization after GPS localization completed
Multiple Slave Card (option, available Q3 2014)
In Master mode enables the synchronization of 2 or more Slaves. Single Slave control and Slave mode supported by mainframe Synchronizationconnector. Specifications to be determined
B3762-2.0 en11HBM
Shipping Case (option, to be ordered separately)
Figure 1.8: Hardcover reusable shipment case with wheels and transport handle
Outside DimensionsWeight Empty CaseSystem Storage AreaAccessories AreaReliable Case TransportCase ExtrasShock and VibrationDrop Test
60.0 cm (23.6\") x 32.5 cm (12.8\") x 67.5 cm (26.5\") (H x W x D)16.5 kg (36.3 lbs)
Special area for system, slides in from the top for easy storage and removal from theshipment case. Protects the system for impact during drop, shock and vibrationsSeparate area for keyboard, mouse, power cable and some cables
Wheels and extendable handle constructed for stable transportation with low gravitationpoint to avoid tumbling of case in any direction during roll transportTwo lift handles and locks on side of the case for easy transportTested with system inside case according to ASTM D4728 ETested with system inside case according to ASTM D4169-04 Level I
Rack Mount Kit (option, to be ordered separately)
Figure 1.9: GEN3i Rack Mount Kit
Rack Mount Kit
Mounting GEN3i mainframe in a standard 19” rack. Does not support the mouse andkeyboard delivered with the GEN3i. Requires no additional mounting materials. Userinstalled option.
HBM12B3762-2.0 en
Txx Adapter (option, to be ordered separately)
Connects a T12 or similar torque transducer to GEN2i, GEN3i and GEN3t Digital Event Timer/Counter connector
Figure 1.10: Txx adapter block diagram
Txx Connection
Number of Txx Torque transducers
Torque interface supportSpeed interface support
Txx signals
LevelsTermination
Differential RS422100 Ω2
Torque and ShuntRPM and Direction
B3762-2.0 en13HBM
Txx Adapter (option, to be ordered separately)
Connects a T12 or similar torque transducer to GEN2i, GEN3i and GEN3t Digital Event Timer/Counter connectorConnectors
GEN series interfaceLoop through event input
Torque interfaceSpeed/RPM interface
Cables
Txx adapter to GEN2i/GEN3i/GEN3t
Txx adapter to Txx TorqueTxx adapter to Txx Speed/RPM
Power
Adapter
Txx Torque transducers
Outside dimensions
Power by GEN2i/GEN3i/GEN3tRequires separate power supply
2.15 (.60”) x 171.6 (6.76”) x 123.6 (4.86) (H x W x D)0.5 m (1. Feet), included with Txx adapter
1-KAB149-6 (other lengths available), to be ordered separately1-KAB163-6 (other lengths available), to be ordered separatelyHD22 sub-D 44 pin male (connecting cable included)
44 pin, female D-type connector, AMP HD-22 series (Tyco: 5748482-5)15 pin, female sub-D type (matches 1-KAB149-6)15pin, female sub-D type (matches 1-KAB163-6)
Figure 1.11: Txx adapter dimensions
HBM14B3762-2.0 en
Acquisition Cards
ModelBasic200kBasic200k XT ISOBasic1MBasic1M ISOBasic1M XT ISOBridge200k ISOBridge1M ISOUni200k ISOUni1M ISOBasic20k-16Basic20k-32HiRes250k-16HiRes250k-32HiSpeed 25MHiSpeed 100MFiber100M 6600Fiber100M 7600Iso1kV200kIso1kV2MMarker1MMarker1M HV(1)(2)(3)(4)
TypeSingle Ended
Unbalanced DifferentialSingle Ended
Unbalanced DifferentialUnbalanced DifferentialBridge/DifferentialBridge/DifferentialDifferential/IEPE/ShuntDifferential/IEPE/ShuntDifferentialDifferential
Differential/IEPE/ChargeDifferential/IEPE/ChargeDifferential/Single EndedDifferential/Single EndedOptical FiberOptical FiberBalanced DifferentialBalanced DifferentialBinaryOptical/Binary
Isolationnoyesnoyesyesyesyesyesyesnonononononoyesyesyesyesnoyes & no
Max. SR(1)200 kS/s200 kS/s1 MS/s1 MS/s1 MS/s200 kS/s1 MS/s200 kS/s1 MS/s20 kS/s20 kS/s250 kS/s250 kS/s25 MS/s100 MS/s100 MS/s100 MS/s200 kS/s2 MS/s1 MS/s1 MS/s
Resolution16 bit16 bit16 bit16 bit16 bit16 bit16 bit16 bit16 bit16 bit16 bit16/24 bit16/24 bit15 bit14 bit--(4)--(4)16/18 bit16/18 bit1 bit1 bit
Memory(2)128 MB128 MB256 MB512 MB512 MB128 MB512 MB128 MB512 MB200 MB200 MB1800 MB1800 MB128 MB1800 MB1800 MB1800 MB200 MB1800 MB512 MB512 MB
Channels88888444416321632444(4)4(4)668 & 32
Event, T/C(3)0, 00, 00, 00, 00, 00, 00, 00, 00, 016, 016, 016, 216, 20, 00, 00, 00, 016, 216, 20, 00, 0
Maximum Sample Rate/channel (not multiplexed).Total recording memory/card.
Digital Events, Timer/Counter channels (Supported by GEN3i and GEN2i Digital Event/Timer/Counter connector only).This card supports maximum four optical fiber transmitter channels.
Optical Fiber Transmitter Channels
Transmitter
Every transmitter is a single channel unit. Every unit has an unbalanced differential input, amplifier, analog anti alias filter and ADC with anoptical data and control link to the receiver card. The receiver card has the recording logic, sample rate selection and memory.ModelHV6600 100MHV6600 25MMV6600 100MMV6600 25M7600 100M
Receiver CardFiber100M 6600Fiber100M 6600Fiber100M 6600Fiber100M 6600Fiber100M 7600
PowerBatteryBattery120/240 V AC120/240 V ACExternal 12 V DC
Sample rate100 MS/s25 MS/s100 MS/s25 MS/s100 MS/s
Resolution14 bit15 bit14 bit15 bit14 bit
Isolation
User application definedUser application defined1800 V RMS1800 V RMS
User application defined
Special Function Cards
5B Carrier
Uses one GEN DAQ slot, holds up to six 5B modules. 5B modules, I/O connectors andcabling not included. An acquisition card is required for actual recording.
The 5B Series signal conditioning modules provide a low cost method of connecting analogsignals with data acquisition systems. They are designed to convert thermocouples, RTD's,strain gages, frequencies, potentiometers, slide wires and other signals into standardized,isolated analog outputs.
B3762-2.0 en15HBM
Ordering Information(1)Article
GEN3i
Description
GEN3i rugged, portable data recorder.(2)
Integrated Instrument, with three acquisitionsslots, 200 MB/s streaming rate, one Master/Slaveconnector, integrated PC, Windows® 7 Ultimate, bit version. 17 inch touch screen TFT,480 GB Solid State Disk, Mouse, Keyboard,carrying bag with integrated protective frontcover. Includes Perception Standard softwarepackage.
GEN3i - 6 channel Iso1kV200k package.
Same as GEN3i plus one 1-GN611-2; 6 channel,1 kV isolated balanced differential inputs,200 kS/s, 128 MB RAM cards.
GEN3i - 12 channel Iso1kV200k package.
Same as GEN3i plus two 1-GN611-2; 6 channel,1 kV isolated balanced differential inputs,200 kS/s, 128 MB RAM cards. (12 channelstotal).
GEN3i - 18 channel Iso1kV200k package.Same as GEN3i plus three 1-GN611-2; 6channel, 1 kV isolated balanced differentialinputs, 200 kS/s, 128 MB RAM cards. (18channels total).
Order No.
1-GEN3i-2
GEN3i plus oneIso1kV200k card
1-GEN3i6-2
GEN3i plus twoIso1kV200k cards
1-GEN3i12-2
GEN3i plus threeIso1kV200k cards
1-GEN3i18-2
(1)(2)
All GEN series systems are intended for exclusive professional and industrial use.
GEN3i ships with keyboard layout in English, German, French, Chinese or Japanese. Specify on order.
HBM16B3762-2.0 en
Options, to be ordered separatelyArticle
IRIG PMC card
Description
Factory installed option.
GEN DAQ IRIG interface fits into open XMC/PMC slot of GEN DAQ Interface/Controllermodule.
Cannot be used in combination with 10 GbitEthernet XMC card.
Factory installed option.
GEN DAQ IRIG/GPS interface fits into openXMC/PMC slot of GEN DAQ Interface/Controllermodule, comes with antenna and 15 m cable.Cannot be used in combination with 10 GbitEthernet XMC card.
GEN DAQ 1GBit Ethernet SFP, 850 nm MultiMode, up to 500 m optical cable lengthsupported, LC connector support.
1 Gbit SFP modules are not compatible with the10 Gbit SFP+ modules.
GEN DAQ 1Gbit Ethernet SFP, 1310 nm SingleMode, up to 10 km optical cable lengthsupported, LC connector support.
1 Gbit SFP modules are not compatible with the10 Gbit SFP+ modules.
Factory installed option.
GEN DAQ 10 Gbit Ethernet XMC card adds upto 2 extra 10 Gbit Ethernet network connectionsto a GEN DAQ series mainframe. Supports up to200 MB/s continuous data transfer from the GENDAQ mainframe to an appropriate PC. Requiresa 10 Gbit optical network SFP+ module. Cannotbe used in combination with IRIG or IRIG/GPSPMC card.
Order No.
1-G001-1Available Q22014
IRIG/GPS PMCcard
1-G002-2Available Q22014
1 Gbit OpticalNetwork SFPmodule 850 nm
1-G062-2
1 Gbit OpticalNetwork SFPmodule 1310 nm
1-G063-2
10 Gbit EthernetXMC card
1-G0-2Available Q22014
10 Gbit OpticalNetwork SFP+module 850 nm
GEN DAQ 10Gbit Ethernet SFP+, 850 nm MultiMode, up to 66 m optical cable length supported,LC connector support.
10 Gbit SFP+ modules are not compatible withthe 1 Gbit SFP modules.
GEN DAQ 10Gbit Ethernet SFP+, 1310 nmSingle Mode, up to 10 km optical cable lengthsupported, LC connector support.
10 Gbit SFP+ modules are not compatible withthe 1 Gbit SFP modules.
1-G065-2Available Q22014
10 Gbit OpticalNetwork SFP+module 1310 nm1-G066-2Available Q22014
B3762-2.0 en17HBM
Accessories, to be ordered separatelyArticle
GEN2i/GEN3i
19 inch rack mountkit
Description
GEN2i/GEN3i rack mount kit (does not includemouse and keyboard mountings).User installed option.
Order No.
1-G053-2
GEN2i/GEN3i/GEN3t shippingcase
GEN2i/GEN3i/GEN3t shipping case, with wheelsand handle Tested according to ASTM D4169-04Level I (drop), and ASTM D4728 E (vibration &shock)
Gross weight (empty) 16.5 kg (36.3 lbs).
1-G0-2
Fiber optic MultiMode standardcable
GEN DAQ standard indoor zipcord fiber opticduplex Multi Mode 50/125 µm cable, 3.0 dB/kmloss, LC-LC connectors, orange, ISO/IEC 11801type OM2.
Used with 850 nm optical 1 Gbit or 10 Gbit
Ethernet. (1-G062-2 and 1-G065-2) and Master/Slave synchronizations.
Lengths 3, 10, 20 and 50 meter (10, 33, 66 and1 feet)
GEN DAQ standard indoor zipcord fiber opticduplex Single Mode 9/125 µm cable, 0.5 dB/kmloss, LC-LC connectors, yellow, ISO/IEC 11801type OS1.
Used with 1310 nm optical 1 Gbit or 10 GbitEthernet (1-G063-2 and 1-G066-2).
Lengths 2, 10, 20, 50 and 100 meter (6.5, 33, 66,1 and 328 feet)
GEN DAQ heavy duty indoor/outdoor fiber opticduplex Single Mode 9/125 µm cable, 0.5 dB/kmloss, LC-LC connectors, black, ISO/IEC 11801type OS1.
Used with 1310 nm optical 1 Gbit or 10 GbitEthernet (1-G063-2 and 1-G066-2).
Lengths 10, 20, 50, 100, 150 and 300 meter (33,66, 1, 328, 492 and 984 feet)
GEN DAQ 5B Carrier. Uses one GEN DAQ slot,holds up to six 5B modules. 5B modules, I/Oconnectors and cabling not included. Basic cardrequired for acquisition.
1-KAB280-31-KAB280-101-KAB280-201-KAB280-50
Fiber optic SingleMode standardcable
1-KAB288-21-KAB288-101-KAB288-201-KAB288-501-KAB288-100
Fiber optic SingleMode heavy dutycable
1-KAB2-101-KAB2-201-KAB2-501-KAB2-1001-KAB2-1501-KAB2-300
5B carrier card1-G028-2
TXX adapter
TXX adapter. Converts the differential signalingused by HBM torque transducers to TTL signallevels used by the timer counter A and B availableon the Digital Event/Timer/Counter connector.Event output connected to Shunt control. Allevent TTL signals available on output connector.Comes with 1 m (3.3 feet) cable to connectadapter to your mainframe. Torque transducercables not included.
1-G070-2
HBM18B3762-2.0 en
Software Options, to be ordered separately(1)ArticleDescription
Order No.
Perception AdvancedSame as Perception Standard but includes the options:
1-PERC-AD-01-2Analysis, Reports, Info Sheets, Exports, Workbooks and Video Playback.Perception Professional
Same as Perception Standard but includes the options
1-PERC-PRO-01-2
Analysis, Reports, Info Sheets, Exports, Workbooks, Video Playback, Basic FFTand Sensor Database.
Perception EnterpriseSame as Perception Professional but runs on native bits Windows® XP, Vista,1-PERC-E-01-2
7 and 8 only. This version supports the use of more than 2 GByte memory inyour PC. Specifically important for systems using multiple mainframes, extensiveuse of the FFT option and/or large amounts of calculations.
Perception Viewer
Same as Perception Standard but without mainframe setup and control.1-PERC-VW-01-2Perception Viewer AdvancedSame as Perception Advanced but without mainframe setup and control.1-PERC-VA-01-2AnalysisAnalysis includes +, -, *, /, max, min, RMS and filters using a formula database.1-PERC-OP-AN-01-2ReportingCreate professional reports including displays, tables, text, graphics.1-PERC-OP-RP-01-2ExportsAdditional export formats added to Perception.
1-PERC-OP-MEX-01-2Info sheetsAllows custom entries and variable definitions to be stored with data.1-PERC-OP-IS-01-2Multiple WorkbooksSimultaneous multiple display windows on multiple monitor systems.1-PERC-OP-MWB-01-2Video PlaybackSimultaneous playback of standard video files.
1-PERC-OP-VP-01-2Basic FFT
Live FFT while recording (hardware dependent) and Review FFT.1 -PERC-OP-BFFT-01Multiple MainframesSimultaneous control of multiple GEN DAQ mainframes.
1-PERC-OP-MMF-01-2Sensor DatabaseCollection of sensors' information, simplifies the set-up of an acquisition channel.1-PERC-OP-SDB-01-2CSI InterfaceAllows the development of and running CSI programs.
1-PERC-OP-CSI-01-2RPC/COMRemote control of Perception, including basic hardware setup and control.1-PERC-OP-IF-01-2BE256/MP controlTo control a single BE256 or a single MultiPro from Perception via IEEE4881-PERC-OP-BE-01-2interface.
SEQUENCETo control BE3200 Test Sequencer from Perception via USB port.1-PERC-OP-SEQ-01-2Requires Perception Standard or higher (1-PERC-ST-01-2).
STL Analysis
Special analysis routines according to the STL standard used in LV, MV and HV1-PERC-OP-STL-01-2
labs. Includes import of TGD data (Test Data Generator) for verification.Requires Analysis option (1-PERC-OP-AN-01-2).
HPHV-AAHighPower/HighVoltage automated analysis. Evaluates data of NoLoad,1-PERC-OP-HHP-01-2
ShortCircuit, Capacitive and Synthetic tests of HV/MV switchgear devices(requires signals from tripping coils and travel to be recorded).Requires STL Analysis option (1-PERC-OP-STL-01-2).
HV-IAHigh Voltage Impulse Analysis option; evaluates Lightning, Switching and1-PERC-OP-HIA-01-2
Current impulses; designed according to IEC60060-1 and IEC61083-2requirements. Allows evaluation with new k-factor method.
eDriveAllows easy and application oriented setup and efficiency calculations of1-PERC-OP-EDR-01-2
electrical inverter / drive tests with minimum interaction.
(1)
Software options are also sold in multiple licenses packages and multiple network license seats.
©Hottinger Baldwin Messtechnik GmbH. All rights reserved.All details describe our products in general form only.
Hottinger Baldwin Messtechnik GmbH
They are not to be understood as express warranty and donot constitute any liability whatsoever.
Im Tiefen See 45 ∙ 293 Darmstadt ∙ GermanyTel. +49 6151 803-0 ∙ Fax: +49 6151 803-9100E-mail: info@hbm.com ∙ www.hbm.com
measure and predict with confidence
ne 0.2-2673B
因篇幅问题不能全部显示,请点此查看更多更全内容
Copyright © 2019- oldu.cn 版权所有 浙ICP备2024123271号-1
违法及侵权请联系:TEL:199 1889 7713 E-MAIL:2724546146@qq.com
本站由北京市万商天勤律师事务所王兴未律师提供法律服务